A Highly-Integrated Analog Front-End IC for Medical Ultrasound Imaging Systems

Translated title of the contribution: 초음파 의료 영상시스템용 고집적 아날로그 Front-End 집적 회로

Research output: Contribution to journalArticlepeer-review

Abstract

A high-voltage highly-integrated analog front-end (AFE) IC for medical ultrasound imaging applications is implemented using standard 0.18-μm CMOS process. The proposed AFE IC is composed of a high-voltage (HV) pulser utilizing stacked transistors generating up to 15 Vp-p pulses at 2.6 MHz, a low-voltage low-noise transimpedance preamplifier, and a HV switch for isolation between the transmit and receive parts. The designed IC consumes less than 0.15 mm² of core area, making it feasible to be applied for multi-array medical ultrasound imaging systems, including portable handheld applications.

Translated title of the contribution초음파 의료 영상시스템용 고집적 아날로그 Front-End 집적 회로
Original languageEnglish
Pages (from-to)49-55
Number of pages7
Journal전자공학회논문지
Volume50
Issue number12
DOIs
StatePublished - 2013

Fingerprint

Dive into the research topics of '초음파 의료 영상시스템용 고집적 아날로그 Front-End 집적 회로'. Together they form a unique fingerprint.

Cite this