An Electrical Stimulator IC with Chopped Pulse based Active Charge Balancing for Neural Interface Applications

Jin Young Son, Hyouk Kyu Cha

Research output: Contribution to journalArticlepeer-review

Abstract

In this paper, a current-mode neural stimulator integrated circuit (IC) using novel active charge balancing technique is presented. The charge balancing technique proposed in this work is based on chopped pulse waveform, where the number of chopped pulses generated in the anodic phase is controlled accurately in order to limit the amount of residual potential at the electrode. In addition, a quick automatic electrode shorting process follows the active charge balancing phase to further discharge to a negligible residual voltage level in every stimulation cycle, ensuring a safe and long-term operation. Both symmetric and asymmetric stimulation pulse waveforms can be selected to provide wide flexibility for various stimulation environment. The stimulator IC designed using 0.18-µm standard CMOS process achieves 12.3 V of voltage compliance and can deliver 1 mA of maximum stimulation current with 5-bit resolution and high linearity. All circuit functions are integrated on-chip without external components, and the fabricated chip consumes only 0.095 mm2 of active die area.

Original languageEnglish
Pages (from-to)322-333
Number of pages12
JournalJournal of Semiconductor Technology and Science
Volume21
Issue number5
DOIs
StatePublished - Oct 2021

Keywords

  • Charge balancing
  • Chopped pulse
  • CMOS
  • Electrical stimulator
  • Neural interface

Fingerprint

Dive into the research topics of 'An Electrical Stimulator IC with Chopped Pulse based Active Charge Balancing for Neural Interface Applications'. Together they form a unique fingerprint.

Cite this