Abstract
Calibration of an X-ray imaging system involves estimation of the 3D geometrical configuration of the system components. Here, we propose a calibration method for stereo X-ray imaging systems, which capture two X-ray images at two orthogonal positions. The calibration parameters include the relation between image coordinates and world coordinates, the relative 3D positions of the X-ray source and detector, and the rotational axis of the stereo X-ray system. The average error of the proposed calibration method was determined to be approximately 0.03% in evaluation tests.
| Original language | English |
|---|---|
| Pages (from-to) | 877-882 |
| Number of pages | 6 |
| Journal | International Journal of Precision Engineering and Manufacturing |
| Volume | 13 |
| Issue number | 6 |
| DOIs | |
| State | Published - Jun 2012 |
Keywords
- Calibration
- Dual X-ray
- Micro-scale X-ray
- Radiographic imaging
- Stereo X-ray