Calibration method for microscale stereo X-ray imaging system

Youngjun Kim, Wontae Kim, Sehyung Park, Jong Hyeong Kim

Research output: Contribution to journalArticlepeer-review

1 Scopus citations

Abstract

Calibration of an X-ray imaging system involves estimation of the 3D geometrical configuration of the system components. Here, we propose a calibration method for stereo X-ray imaging systems, which capture two X-ray images at two orthogonal positions. The calibration parameters include the relation between image coordinates and world coordinates, the relative 3D positions of the X-ray source and detector, and the rotational axis of the stereo X-ray system. The average error of the proposed calibration method was determined to be approximately 0.03% in evaluation tests.

Original languageEnglish
Pages (from-to)877-882
Number of pages6
JournalInternational Journal of Precision Engineering and Manufacturing
Volume13
Issue number6
DOIs
StatePublished - Jun 2012

Keywords

  • Calibration
  • Dual X-ray
  • Micro-scale X-ray
  • Radiographic imaging
  • Stereo X-ray

Fingerprint

Dive into the research topics of 'Calibration method for microscale stereo X-ray imaging system'. Together they form a unique fingerprint.