Abstract
Optical pick-up is a core component for data read/write operations in optical disc drives, and an optical pick-up performance evaluator is an instrument used to analyze the overall performance of an optical pick-up. Due to inevitable errors in an analog measurement circuit, resultant evaluation data is not guaranteed to be accurate. In this paper, a calibration method for an optical pick-up performance evaluator is proposed to ensure evaluation accuracy. Measured data is corrected by a 1st order correction function, and a calibration process based on least-square method is utilized to obtain correction coefficients of the correction function. The proposed calibration method is applied to experiments, and enhanced accuracy is presented with resultant evaluation data.
| Original language | English |
|---|---|
| Pages (from-to) | 578-283 |
| Number of pages | 296 |
| Journal | Journal of Institute of Control, Robotics and Systems |
| Volume | 20 |
| Issue number | 5 |
| DOIs | |
| State | Published - 2014 |
Keywords
- Absolute accuracy
- Calibration
- Least-square method
- Optical pick-up
- Performance evaluator