TY - JOUR
T1 - Quality factor measurement of micro gyroscope structure according to vacuum level and desired Q-factor range package method
AU - Kim, Jong Seok
AU - Lee, Sang Woo
AU - Jung, Kyu Dong
AU - Kim, Woon Bae
AU - Choa, Sung Hoon
AU - Ju, Byeong Kwon
PY - 2008/6
Y1 - 2008/6
N2 - A micro-machined gyro chip of gyroscope is normally packaged in specific vacuum level to get the specific quality factor(Q-factor). If the Q-factor is too high, frequency tuning and the approximate matching between driving and sensing comb structure become difficult, and if the Q-factor is too low, its sensitivity decreases. The optimum Q-factor of our gyro chip design is 4000 range. To get this range, we measured the drive mode Q-factor as vacuum level of our gyro chip and we found that the vacuum level of the desired Q-factor 4000 is in the range of 740 mTorr. Based on this data, we fabricate the wafer level package gyro chip of the desired Q-factor by controlled the basic pressure of package bonding chamber just prior to the bonding process. After wafer level package process, we measured Q-factor of whole samples. Among 804 samples, 502 packaged gyro chips are worked and the Q-factor of 67% samples is between 3500 and 4500 range.
AB - A micro-machined gyro chip of gyroscope is normally packaged in specific vacuum level to get the specific quality factor(Q-factor). If the Q-factor is too high, frequency tuning and the approximate matching between driving and sensing comb structure become difficult, and if the Q-factor is too low, its sensitivity decreases. The optimum Q-factor of our gyro chip design is 4000 range. To get this range, we measured the drive mode Q-factor as vacuum level of our gyro chip and we found that the vacuum level of the desired Q-factor 4000 is in the range of 740 mTorr. Based on this data, we fabricate the wafer level package gyro chip of the desired Q-factor by controlled the basic pressure of package bonding chamber just prior to the bonding process. After wafer level package process, we measured Q-factor of whole samples. Among 804 samples, 502 packaged gyro chips are worked and the Q-factor of 67% samples is between 3500 and 4500 range.
UR - http://www.scopus.com/inward/record.url?scp=45849146963&partnerID=8YFLogxK
U2 - 10.1016/j.microrel.2008.03.001
DO - 10.1016/j.microrel.2008.03.001
M3 - Article
AN - SCOPUS:45849146963
SN - 0026-2714
VL - 48
SP - 948
EP - 952
JO - Microelectronics Reliability
JF - Microelectronics Reliability
IS - 6
ER -